![background image](http://html2.mh-extra.com/html/racal-instruments/3151/3151_user-manual_3124279035.webp)
3151 And 3151A User Manual
Using The Instrument 3-5
<delay>,<high_time>,<rise_time>,<fall_time>}
programs the generator to output a pulse waveform with frequency,
amplitude, offset, delay, rise time, high time and fall time parameters.
The default settings for these functions are: 1 MHz, 5 Vp-p, 0 V, 0%,
10%, 10% and 10%.
The command:
APPLy:RAMP {<frequency>,<amplitude>,<offset>,
<delay>, <rise_time>,<fall_time>}
programs the generator to output a ramp waveform with frequency,
amplitude, offset, delay, rise time, and fall time parameters. The
default settings for these functions are: 1 MHz, 5 Vp-p, 0 V, 0%, 10%
and 10%.
The command:
APPLy:SINC {<frequency>,<amplitude>,<offset>,
<number_cycles>}
programs the generator to output a sine(x)/x waveform with
frequency, amplitude, offset, and number of cycles parameters. The
default settings for these functions are: 1 MHz, 5 Vp-p, 0 V and 10.
The command:
APPLy:EXPonential <frequency>,<amplitude>,<offset>,
<exponent>}
programs the generator to output an exponential waveform with
frequency, amplitude, offset, and exponent parameters. The default
settings for these functions are: 1 MHz, 5 Vp-p, 0 V and -10.
The command:
APPLy:GAUSsian {<frequency>,<amplitude>,<offset>,
<exponent>}
programs the generator to output a gaussian waveform with
frequency, amplitude, offset, and exponent parameters. The default
settings for these functions are: 1 MHz, 5 Vp-p, 0 V and 10.
The command:
APPLy:DC {<percent_amplitude>}
programs the generator to output a DC level. The DC level is set as a
percent of programmed amplitude. The default setting for this
function is 100%.
Summary of Contents for 3151
Page 4: ...This page was left intentionally blank...
Page 24: ...3151 And 3151A User Manual Getting Started 1 12 This page was left intentionally blank...
Page 29: ...3151 And 3151A User Manual Configuring The Instrument 2 5 Figure 2 1 Set The Logical Address...
Page 74: ...3151 And 3151A User Manual Using The Instrument 3 44 This page was left intentionally blank...
Page 98: ...3151 And 3151A User Manual SCPI Command Reference 4 24 This page was left intentionally blank...
Page 127: ...3151 And 3151A User Manual Optional Harness Assemblies 6 3...
Page 131: ...3151 And 3151A User Manual Optional Harness Assemblies 6 7...