1261B User Manual
Racal Instruments
Ó
1995
Using The Enhanced Monitoring System 4-35
Programming The
Number of Data
Bits
The number of data bits may be programmed with the
command:
SYSTEM:COMM:SERIAL:BITS <data bits>
where “<data bits>” will either be 7 or 8. The presently
programmed number of data bits may be read with the query:
SYSTEM:COMM:SERIAL:BITS?
Programming The
Parity
The parity may be programmed using one of the following
commands:
SYSTEM:COMM:SERIAL:PARITY EVEN
SYSTEM:COMM:SERIAL:PARITY
ODD
SYSTEM:COMM:SERIAL:PARITY
NONE
The present parity may be read using the query:
SYSTEM:COMM:SERIAL:PARITY?
The EMS will reply to this query with one of the following:
EVEN
ODD
NONE
Programming The
Number of Stop
Bits
The number of data bits may be programmed with the
command:
SYSTEM:COMM:SERIAL:SBITS <stop bits>
where “<stop bits>” will either be 1 or 2. The presently
programmed number of stop bits may be read with the query:
SYSTEM:COMM:SERIAL:SBITS?