1261B User Manual
Using The Enhanced Monitoring System 4-38
Racal Instruments
Ó
1995
Reading Event
Time Stamps
Each alarm condition has an associated time stamp. This time
stamp tracks the most recent occurrence of the out-of-tolerance
condition. The time stamp indicates the earliest time at which
the most recent out-of-tolerance condition was detected.
In general, the time stamps may be read using a query of the
form:
<attribute>:ALARM:TIME?
-- read the time of the event
<attribute>:ALARM:DATE?
-- read the date of the event
Where <attribute> is replaced by a keyword which indicates
which attribute’s out-of-tolerance time stamp is to be read. The
valid queries are:
VOLT1:ALARM:TIME?
-- +5V PS voltage alarm
VOLT2:ALARM:TIME?
-- -5.2V PS voltage alarm
VOLT3:ALARM:TIME?
-- -2V PS voltage alarm
VOLT4:ALARM:TIME?
-- +24V PS voltage alarm
VOLT5:ALARM:TIME?
-- -24V PS voltage alarm
VOLT6:ALARM:TIME?
-- +12V PS voltage alarm
VOLT7:ALARM:TIME?
-- -12V PS voltage alarm
CURR1:ALARM:TIME?
-- +5V PS current alarm
CURR2:ALARM:TIME?
-- -5.2V PS current alarm
CURR3:ALARM:TIME?
-- -2V PS current alarm
CURR4:ALARM:TIME?
-- +24V PS current alarm
CURR5:ALARM:TIME?
-- -24V PS current alarm
CURR6:ALARM:TIME?
-- +12V PS current alarm
CURR7:ALARM:TIME?
-- -12V PS current alarm
FREQ1:ALARM:TIME?
-- most recent alarm of any fan
FREQ2:ALARM:TIME?
-- fan #1 alarm
FREQ3:ALARM:TIME?
-- fan #2 alarm
FREQ4:ALARM:TIME?
-- fan #3 alarm
TIME1:ALARM:TIME?
-- power on time alarm
TIME2:ALARM:TIME?
-- cumulative power on time
TIME3:ALARM:TIME?
-- filter service time alarm
TEMP1:ALARM:TIME?
-- slot 0 temp rise alarm
TEMP2:ALARM:TIME?
-- slot 1 temp rise alarm
TEMP3:ALARM:TIME?
-- slot 2 temp rise alarm
TEMP4:ALARM:TIME?
-- slot 3 temp rise alarm
TEMP5:ALARM:TIME?
-- slot 4 temp rise alarm
TEMP6:ALARM:TIME?
-- slot 5 temp rise alarm
TEMP7:ALARM:TIME?
-- slot 6 temp rise alarm
TEMP8:ALARM:TIME?
-- slot 7 temp rise alarm
TEMP9:ALARM:TIME?
-- slot 8 temp rise alarm
TEMP10:ALARM:TIME?
-- slot 9 temp rise alarm
TEMP11:ALARM:TIME?
-- slot 10 temp rise alarm
TEMP12:ALARM:TIME?
-- slot 11 temp rise alarm
TEMP13:ALARM:TIME?
-- slot 12 temp rise alarm
TEMP14:ALARM:TIME?
-- ambient temperature alarm
VXI:SYSFAIL:ALARM:TIME?
-- VXI SYSFAIL* alarm