1261B User Manual
Error Messages B-4
Racal Instruments
Ó
1995
-311,”Memory error”
DDE
The EMS’s memory is corrupted.
-314,”Save/recall memory lost”
DDE
The *SAV or *RCL command has failed due to a corrupted
nonvolatile memory.
-315,”Configuration memory lost”
DDE
Configuration information has been lost due to a corrupted
nonvolatile memory.
-330,”Self-test failed”
DDE
The self-test has failed. If this is set at power-up, execute the
*TST? query to determine what has failed.
-350,”Queue overflow”
DDE
An attempt was made to add an error message to a full error
queue. Thus, the error queue has overflowed.
-360,”Communication error”
DDE
A communication error has occurred over the RS-232
interface.
-361,”Parity error in program message”
DDE
A parity error has been detected in the command over the RS-
232 interface.
-362,”Framing error in program message”
DDE
A framing error has been detected in the command over the
RS-232 interface.
-363,”Input buffer overrun”
DDE
The RS-232 interface input buffer has been overrun.
-400,”Query error”
QYE
A query error has been detected.
-410,”Query INTERRUPTED”
QYE
The query has been interrupted (see IEEE-488.2, Sec.
6.3.2.3).
-420,”Query UNTERMINATED”
QYE
see IEEE-488.2, Sec 6.3.2.2 for a description.
-430,”Query DEADLOCKED”
QYE
The query has reached a deadlocked state. This happens
when the input buffer is full, the output buffer is full, and a reply
to a query cannot be generated.
1000,”VXI signal queue overflow”
DDE
The signal queue has overflowed. Sending VXI signals is an
alternative method of generating interrupts.
1001,”VXI interrupt queue overflow”
DDE
The queue of pending interrupts to send has overflowed. The
EMS maintains a queue of interrupts which should be sent.
An interrupt cannot be sent until a previous interrupt has been
acknowledged.
1002,”VXI bus error”
DDE
A VXIbus error has occurred when the EMS attempted to write
data over the VXIbus.
1003,”VXI bus timeout”
DDE
A timeout has occurred while the EMS was waiting for control
of the VXIbus.
1005 - 1999, “Internal error: xxx”
DDE
An internal firmware error has been detected. Errors of this
type should be reported to Racal Instruments for correction.