1261B User Manual
Configuring The 1261B 2-18
Racal Instruments
Ó
1995
Enhanced
Monitor
Configuration
Switch
Omit this procedure if your mainframe is not equipped with the
Enhanced Monitoring System (EMS).
NOTE
Perform this procedure only if factory default setting
is incompatible with your application. See Figure 2-12
for default setting.
WARNING
Ensure that J400 AC Input is disconnected before
removing or replacing the System Monitoring (EMS)
module. Possible damage to the module could result.
1. Remove the EMS module. See Removal
and Replacement in Chapter 7.
Position “1” enables changes.
2.
Configure SW1 located on the
component side of the EMS module.
3. Replace the EMS module. See Removal
and Replacement in Chapter 7
Figure 2-12, Setting the Enhanced Monitoring System Configuration Switch
Revised 5/21/01