MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
292
System Exception Error
Manual Restart Required
* Manually Restart System
* Call service after multiple
occurrences
293
EEPROM Reset to Defaults. Please
Restart
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
alarms
295
System settings Data Error
Settings Reset to Defaults
CONFIRM
* Continue as normal
* Check Settings in Menu
296
Case Data Error
Reset to Defaults
CONFIRM
* Continue as normal. Case
Data
is not essential to operation.
* Call service after multiple
occurrences
297
Delivered Volume Data Error
Reset to Defaults
CONFIRM
* Continue as normal. Check
delivered volumes for
accuracy
* Call service after multiple
occurrences
298
Master file Unavailable
Please insert media with Master File
CONFIRM
* Master file may be corrupt
* Re-download Master File
* Call service after multiple
occurrences
299
Incompatible Master file detected
Please insert media with compatible file
CONFIRM
* Master file may be corrupt
* Re-download Master File
* Call service after multiple
occurrences
300
Customized data corrupted
All customized data needs review
CONFIRM
* All Personnel, Additive,
Crystalloid, and Component
data should be reviewed
301
CaseLog Version Mismatch/Recovery Error
CONFIRM
* Retry
* Try another Flash memory
device
302, 303
System maintenance due
Please schedule System Maintenance
CONFIRM
* Continue as normal
* Contact service to schedule
maintenance
304, 305
Audio module Error
Check the screen for all
Alarms/Notifications
CONFIRM
* Audio tones may not be heard
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
38