MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
59
Delivery occlusion
Check for clamps
Check pressure limit setting
CONFIRM
* Check delivery line for
clamps,
kinks, or occlusions
* Check the upper pressure
limit
settings
60
System Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
61, 62
Inadequate Crystalloid fill
Check Source
CONFIRM
* Check Crystalloid inlet line for
clamps, kinks, or occlusions
* Raise the Crystalloid bag or
use
a pressure cuff
63, 64
Inadequate Blood fill
Check Source
CONFIRM
* Check Blood inlet line for
clamps, kinks, or occlusions
* Increase the blood inlet
source
pressure
65
Inadequate Fill Check Source
CONFIRM
* Check blood/crystalloid
source
lines and pressures
* If problem persists, restart
system
* Call service after multiple
occurrences
66, 67
System Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
68, 69
System FPGA Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
70
System Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
71, 72
System FPGA Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
73
System Mechanical Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
13