MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
134, 135,
136
Heater diagnostics failed
Continue with heaters disabled
Start a NEW CASE to retest heaters
CONFIRM
* Continue if heating is not
required for procedure
* Ensure the H2O Circulation
System has been fully primed
using the steps outlined in the
Operations manual
* Restart and select New Case
to
retest heater diagnostics
* Contact service after multiple
occurrences.
137
Heater diagnostics started
May take up to 100 seconds
CONFIRM
* Wait for heater diagnostics to
complete
138
Pneumatic pressure sensor error
Service required
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
139
Low Pneumatic pressure detected
SHUTDOWN
RESTART
* Connect to medical air
* Call service following
procedure
140
Blood/Cryst pump has been stopped
Pneumatic pressure critical
Check compressor or connect medical
air
SHUTDOWN
RESTART
* Connect to medical air
* Call service following
procedure
141
Pneumatic overpressure (> 55 psi)
Check medical air source
SHUTDOWN
RETRY
* Disconnect from medical air
and
retry
Alar
m
Cod
e
Message Display
Corrective Action
23