MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
164
Additive Cartridge not detected
Verify Additive cartridge installation
CONFIRM
* Check Additive Cartridge to
verify it is properly installed
and
fully secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
system and try again
* Contact service after multiple
occurrences
165
Elevated internal temperature detected
Check for obstructed air flow
Maintenance required
CONFIRM
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
166
Internal temperature Error detected
Check for obstructed air flow
Unable to proceed. Call for Service
SHUTDOWN
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
167
Elevated internal temperature detected
Check for obstructed air flow
Maintenance required
CONFIRM
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
168
Internal temperature Error detected
Check for obstructed air flow
Unable to proceed. Call for Service
SHUTDOWN
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action