MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
209, 210,
211, 212,
213, 214,
215, 216,
217, 218,
219, 220
Data Logging Error
Consult Service after this case
CONFIRM
* Continue as normal
* Contact service after multiple
alarms
221, 222,
223
Data Log File Transfer
CONFIRM
* Continue as normal
* Contact service after multiple
alarms
224
Arrest pump error
Arrest pump is Off
CONFIRM
* Check Arrest
Cartridge to
verify it is properly installed
and
fully secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
console and try again
* Contact service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
34