MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
150
Arrest Cartridge not detected
Verify Arrest cartridge installation
CONFIRM
* Check Arrest Cartridge to
verify
it is properly installed and fully
secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
system and try again
* Contact service after multiple
occurrences
151
H2O temperature sensor error
H2O temperature will not be displayed
Heaters Disabled
CONFIRM
* Continue without temperature
control
OR
* Restart and resume case
* Call service after multiple
occurrences
152
Pneumatic pressure sensor error
Service required
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
153
Heater controls failed
Continue with heaters disabled
Start a NEW CASE to retest
CONFIRM
* Continue without temperature
control
OR
* Restart and start New Case
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action