MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
39
System pressure sensor error
RETEST
* Check the lines for clamps,
kinks or occlusions
* Manually actuate delivery and
vent valves several times
* Check heat exchanger to
ensure it is securely
* Restart System
* Contact service after multiple
occurrences
40
External Antegrade sensor not detected
USE SYSTEM
RETEST
* Select USE SYSTEM
OR
* Retest and if multiple failures
are observed, disconnect the
cable and try again
* Contact service after multiple
occurrences
41
External Retrograde sensor not detected
USE SYSTEM
RETEST
* Select USE SYSTEM
OR
* Retest and if multiple failures
are observed, disconnect the
cable and try again
* Contact service after multiple
occurrences
42
Max System pressure
Check circuit
CONFIRM
* Check the lines for clamps,
kinks or occlusions
* Manually actuate delivery and
vent valves several times
* Check heat exchanger to
ensure it is tightly secured
with
the locking knob
* Restart System
* Contact service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action