MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
74
Door is open
Close door to continue
SHUTDOWN
* Close the door
* Press shutdown to turn
System
off
75, 76
System FPGA Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
78, 79,
84, 85
System Error
SHUTDOWN
RESTART
* Restart System
* Call service after multiple
occurrences
86, 87, 88,
89
Blood Pump Error
Flow may be resumed
SHUTDOWN
CONFIRM
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
92, 93
Unable to maintain flow rate setting
Check circuit
CONFIRM
* Check blood/crystalloid
source
lines and pressures
* Stop flow and restart flow
* If problem persists, restart
system
* Call service after multiple
occurrences
94
Mechanism valve error
Reinitializing. Please wait…
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
95
Delivery valve sensor mismatch
Flow may be resumed
CONFIRM
* Continue with procedure
* If error occurs multiple times
over the course of the
procedure, contact service
immediately following
procedure
96, 97, 98
Arrest pump failed diagnostics
Proceed without Arrest pump
RESTART
CONFIRM
* Restart System
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
14