Troubleshooting
QIAcube User Manual 06/2008
7-15
To continue sample processing:
1.
Note the step at which the protocol stopped. This is
displayed in the touchscreen.
2.
Remove the samples and reagents from the QIAcube.
3.
Refer to the appropriate protocol in the relevant kit
handbook and continue sample processing manually.
7.2.7
Centrifuge
Opening the centrifuge lid in the event of a
breakdown
In case of power failure, the centrifuge lid can be manually
opened so that the samples can be removed. To open the
centrifuge lid, follow the instructions below.
WARNING
Moving parts
[W12]
In case of breakdown caused by power failure, remove the
power cord and wait 10 minutes before attempting to
manually open the centrifuge lid.