Troubleshooting
QIAcube User Manual 06/2008
7-5
Error
code
Description
Comments and suggestions
405 Protocol error: unable to
save protocol
Delete protocols from the QIAcube that
are no longer required and reload the
protocols from the USB stick (see
Section 5.4).
Switch off the QIAcube, wait for a few
minutes, and switch it back on again. If
the error persists, contact QIAGEN
Technical Services.
406 Protocol
error:
maximum
number of protocols
Delete protocols from the QIAcube that
are no longer required (see Section
5.4).
Switch off the QIAcube, wait for a few
minutes, and switch it back on again. If
the error persists, contact QIAGEN
Technical Services.
407 Protocol error: unable to
save protocol list
Delete protocols from the QIAcube that
are no longer required (see Section
5.4).
Switch off the QIAcube, wait for a few
minutes, and switch it back on again. If
the error persists, contact QIAGEN
Technical Services.
408 Protocol
error:
max.
protocol steps exceeded
Contact QIAGEN Technical Services for
a new protocol.
409 Protocol
error:
missing
liquid definition
Contact QIAGEN Technical Services for
a new protocol.
410 Protocol
error:
missing
vessel definition
Contact QIAGEN Technical Services for
a new protocol.
412 Protocol
error:
missing
working space definition
Contact QIAGEN Technical Services for
a new protocol.
Summary of Contents for QIAcube
Page 1: ...Version 1 1 June 2008 Sample Assay Technologies QIAcube User Manual...
Page 40: ...Introduction 2 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 43: ...General Description QIAcube User Manual 06 2008 3 3...
Page 118: ...Maintenance Procedures 6 20 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 140: ...Troubleshooting 7 22 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 150: ...Appendix B B 2 QIAcube User Manual 06 2008 Centrifuge Shaker...
Page 151: ...Appendix B QIAcube User Manual 06 2008 B 3 Centrifuge Shaker...
Page 152: ...Appendix B B 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 156: ...Appendix C C 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 158: ...Appendix D D 2 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 161: ......