Installation Procedures
4-2
QIAcube User Manual 06/2008
4.2
AC power connection
Power requirements
The QIAcube operates at:
100–120 V AC, 50/60 Hz, 650 VA (North America and
Japan)
220–240 V AC, 50/60 Hz, 650 VA (Europe)
Make sure that the voltage rating of the QIAcube is
compatible with the AC voltage available at the installation
site. Mains supply voltage fluctuations are not to exceed 10%
of nominal supply voltages.
Grounding requirements
To protect operating personnel, the National Electrical
Manufacturers’ Association (NEMA) recommends that the
QIAcube be correctly grounded (earthed). The instrument is
equipped with a 3-conductor AC power cord that, when
connected to an appropriate AC power outlet, grounds
(earths) the instrument. To preserve this protection feature,
do not operate the instrument from an AC power outlet that
has no ground (earth) connection.
Installation of AC power cord
Connect one end of the AC power cord to the socket located
at the rear of the QIAcube, and the other end to the AC
power outlet.
4.3
Unpacking the QIAcube
WARNING
Risk of personal injury and material damage
[W2]
The QIAcube is too heavy to be lifted by one person. To
avoid personal injury or damage to the instrument, do not
lift the instrument alone.
Summary of Contents for QIAcube
Page 1: ...Version 1 1 June 2008 Sample Assay Technologies QIAcube User Manual...
Page 40: ...Introduction 2 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 43: ...General Description QIAcube User Manual 06 2008 3 3...
Page 118: ...Maintenance Procedures 6 20 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 140: ...Troubleshooting 7 22 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 150: ...Appendix B B 2 QIAcube User Manual 06 2008 Centrifuge Shaker...
Page 151: ...Appendix B QIAcube User Manual 06 2008 B 3 Centrifuge Shaker...
Page 152: ...Appendix B B 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 156: ...Appendix C C 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 158: ...Appendix D D 2 QIAcube User Manual 06 2008 Page left intentionally blank...
Page 161: ......