Tranducer detection and Hardware settling
In this menu, the transducer detection feature and the settling of the
amplifier stages in VIBXPERT can be switched on and off. Additional
information can be found in the 'Route Setup' section (p. 2-17).
The settling of the amplifier can only be activated when the trans-
ducer detection is switched off.
The settings in this menu apply to the 'Multimode' operating mode
only. The settings for the 'Route' and 'Machine template' modes are
made in the ‘Route Setup’ menu.
After the device is switched on, the amplifiers always settle prior to
the first measurement.
Sensor test
Here you can manually check the measurement path to the sensor. For
ICP sensors, use the 'ICP Sensor Test' menu; for all other sensors, use
the 'Sensor Test' menu.
• Select the channel to which the sensor is connected (A, B).
• Click on START.
VIBXPERT then checks the measurement path and displays the re-
sult ('LineDrive', 'Open line', 'Short circuit', etc.).
Default vibration sensor (Multimode)
A default sensor can be specified for all vibration measurements in the
'Multimode' mode. In the route / machine template, a default sensor
can only be set up via the 'Runtime setup' menu (p. 3-18) or the Route
Device Setup (p. 2-17) respectively.
• Activate the checkbox and select the vibration sensor.
Note
Getting started - Basic settings
Left:
Checking the measurement
route on channel B
Right:
Standard sensor for vibration mea-
surement in 'Multimode'