NOTICE!
Clean the lamp protection tubes each time a lamp
is replaced.
Dispose of used lamps in accordance with the
applicable guidelines and directives. Usually these
can be disposed of together with used fluorescent
tubes.
Safety switch for the UV lamp cover
The UV lamp cover is protected with a safety
switch. The safety switch has a reed switch and
comprises two components. The safely switch
ensures that the UV lamp only generates UV radia‐
tion in its fitted state. If the safety switch is
removed, then the UV lamp switches off. If the
safety switch is not fitted and thus remains closed,
e.g. after installation and maintenance work, then
the UV lamp cannot be started.
1.
Switch off the UV system
2.
Switch off the main switch or disconnect it from the mains
power supply
3.
Close the shut-off valves upstream and downstream of the
radiation chamber
4.
Remove the safety switch of the UV lamp cover
5.
Depending on the system size, now first remove the wiper
motor
6.
Loosen the fixing bolts of the lamp cover using an Allen key,
remove the lamp cover and pull out the lamp by approxi‐
mately 100 mm
7.
Loosen the UV lamp at the cable clamp with the aid of a
slotted head screw driver
8.
Fully remove the UV lamp and place it to one side
9.
Check whether the O-ring on the UV lamp protection tube
bracket is in the groove provided - ensure that the sealing
surfaces of the O-ring are completely smooth and clean!
10.
Position of the lamp return cable.
Check that the lamp return cable is not lying
between the UV lamp and the UV sensor.
The shadow of the lamp return cable could
otherwise distort the measurement of the UV
intensity.
Insert the new UV lamp into the UV lamp protection tube and
allow it to project out approx. 100 mm
11. Fix the UV lamp to the cable clamp with the aid of a slotted
head screw driver
12. Insert the UV lamp fully into the UV lamp protection tube
13. Place the UV lamp cover onto the UV lamp protection tube
bracket and, using the fixing bolts provided, screw in and
tighten with an Allen key
14. Fit the safety switch of the UV lamp cover.
15. Depending on the system size, now fit the wiper motor
Maintenance