AMMONIA User Manual
AMMONIA_User 9/09
6
0
6.2 Specifications
Communications
User Interface
Rating - Detail
Notes
Keypad - LCD
5 Key Tactile feedback: UP /
DOWN / ENTER / EXIT / RIGHT
2 Line x 16 Character, Backlit
Scan rate 100mS nominal
User adjustable contrast
USB
Data log download.
View-configuration upload.
Includes display support for
USB connection state
10 BaseT, TCP-IP
Ethernet LAN
(Optional)
HTML – AJAX micro Web Server
for Mozillla’s Firefox and IE7
Full command, control, reconfigure
via browser.
Network parameters and HTML
port user set. Static IP
Password, UserID protected.
Auto-configures views linking
sensors and controls.
Modem
(Optional)
56K, V.90 PPP access
provides remote browsing via
Mozillla’s Firefox and IE7
Alarms accessed via Trackster3
polling.
Electrical
Rating - Detail
Notes
AC Input
120 or 230 VAC, 50/60Hz,
Switch selectable
5 Relay Fusing
6.3 Amps @ 120VAC
3.15 Amps @ 230VAC
5x20mm, AC fusing:
Relays alarm & OFF on fusing.
120VAC Power
Cord
0 to 1, AWG18, NEMA grounded
plug, SJTW jacket, black, 60”
Quantity varies with Analyzer
part number.
120VAC Pump-Solenoid
Plug Sockets
0 to 5, AWG18, 3 wire, 12” long
NEMA receptacle sockets, black.
Quantity varies with Analyzer
part number.
Surge-Spike Suppression
Control Fusing
Relays 2-5 NO contacts
snubbed 0.1uF, 150R
Varistor on control AC input
Analyzer, transformer isolated
from AC line.
Control fusing not user
accessible.
AC Terminals
AWG 14, 150mm
2
UL rated 300V @ 10A
Rising clamp type.
Electrical grounds at bottom of
aluminum backplate.
Conductor insulation rated
600VAC minimum.
Sensor, Digital Input
Terminals
AWG 22, 0.25 – 0.50mm
2
MAX AWG14, 150 150mm
2
DC Loop – Turbine Meter
Power
15 – 22 VDC, unregulated
Thermally fused @ 100mA hold,
trip @ 200mA
Available at 3 field wiring
terminals:
+DC Power Output
APPENDIX A