AMMONIA User Manual
AMMONIA_User 9/09
3
0
2.3 Sensor Alarms
To adjust a sensor alarm, key
UP
or
DOWN
to the target sensor and press
ENTER
.
Key
UP
and
ENTER
@
Alarms
.
Key
UP
or
DOWN
to select
High Alarm
,
Low Alarm
or
Delay on Alarm
& press
ENTER
.
Key
UP
or
DOWN
to change the underlined digit.
Key
RIGHT
to move the digit underline.
Press
ENTER
to modify.
or
EXIT
to leave unchanged.
In this example we’ve increased the
High Alarm
from
1.60 ppm
to
1.90 ppm
.
NH3
1.30 ppm
Alarms
Configure
High Alarm
1.60 ppm
D
High Alarm
1.90 ppm
Adjust Alarms
and
Editing, or Exit
1.90 ppm
then
Sidebar:
Reset Alarms:
Section 1.2
Clear Alarms
resets the
Delay on Alarm
time
If the
Delay on Alarm
is set to zero minutes and the sensor is above the
High Alarm
or below the
Low Alarm
, the sensor alarm will immediately re-trip.