AMMONIA User Manual
AMMONIA_User 9/09
5
6
5.7 Communications
This display sequence is only applicable to
Analyzers with the ‘
LB
’, LAN-Browser option.
To view or adjust the Ethernet
IP Address
press
ENTER
and
DOWN
to
Communicate
at the power up or day-time display.
Key
ENTER
@
Communicate
Key
ENTER
to modify.
Key
UP
or
DOWN
to change the underlined number
Key
RIGHT
to move to the next 3 digit number.
Key
ENTER
to change or
EXIT
to leave
the
IP Address
unchanged.
In this example we’ve changed the
IP Address
from
10.10.6.106
to
192.168.24.86
.
Modify IP Address
Thu 16:54:10
S/N: A076X486
Communicate
Configure
then
IP Address
10.10.6.106
IP Address
192.168.24.86
IP Address
192.168.024.106
then
Sidebar:
Warning:
Do not connect the Analyzer Ethernet connection into a site LAN
until the site IT staff have assigned a valid IP Address.
Using An Ethernet CrossOver Cable:
You’ll need to set your notebook PC’s IP Address to the same network to browse the Analyzer
using a crossover cable. In this pages example, the Analyzer
IP Address
is
192.168.24.86
.
To be on the same network, your notebook needs an
IP Address
192.168.24.
xxx
where xxx is any number from 1 to 255, excluding
86
.