AMMONIA User Manual
AMMONIA_User 9/09
5
5
5.7 Communications
This display sequence is only applicable to
Analyzers with the ‘
LB
’, LAN-Browser option.
See appendix LB
To view or adjust the Analyzer Ethernet setting
press
ENTER
and
DOWN
to
Communicate
at the power up or day-time display.
Key
ENTER
@
Communicate
Displays the current LAN
IP address
.
In this example, it’s the factory default.
If you are connecting into the site LAN,
IP address
is assigned by the site IT staff.
Key
ENTER
to modify.
Netmask
is usually this value for most sites.
Key
ENTER
to modify.
This is the default
HTML Port
for browsing.
It can be modified only via the browser.
The site IT staff may require the Analyzer
MAC Address
to
register-allow the Analyzer on the site network.
The
MAC address
is six 2 digit hexadecimal
numbers, separated by colons into
3 groups of 4 to fit the LCD screen.
In this example, the
MAC address
is
00 90 C2 00 00 00
Communicate
Thu 16:54:10
S/N: A076X486
Communicate
Configure
Netmask
255.255.255.0
HTML Port
80
MAC Address
0090:c200:0000
then
or
or
IP Address
10.10.6.106
IP Address
10.10.6.106
Sidebar:
Warning:
Do not connect the Analyzer Ethernet connection into a site LAN
until the site IT staff have assigned a valid IP Address.
Not Connected to the Site LAN:
Leave the IP Address at 10.10.6.106. Connect a crossover cable from your notebook PC to the
Analyzer and browse 10.10.6.106.
Browser passwords are the same as the default keypad passwords listed
in manual Section
5.1 Passwords
.
You’ll need to configure your notebook to connect.
Refer to Appendix ‘
C
’
‘LB’ LAN Browser Option
at the back of this manual.