AMMONIA User Manual
AMMONIA_User 9/09
5
4
5.6 Configurations
To view or adjust the configuration of the whole Analyzer
press
ENTER
and
DOWN
to
Configure
at the power up or day-time display.
Key
ENTER
@
Configure
Requires ‘LB’ Option
Load Factory
configures the Analyzer to one of the 1 to 15
configurations shipped with the Analyzer.
This Analyzer is currently running the AMMONIA
configuration.
. View and select an alternative Analyzer configuration
by keying
ENTER
.
Load Config.
returns to the last saved configuration.
Key
ENTER
,select
Yes
&
ENTER
.
After you’ve modified the Analyzer for your site and
its water treatment program,
Save config.
by
keying
ENTER
& selecting
Yes
Once saved, you can always recover by using
Load config
.
The
A&B
Analyzer inputs may be configured for several
sensor types. Usually this is done once/site.
Key
ENTER
to view or modify the current
A&B
Sensor
type.
Thu 16:54:10
S/N: A076X0486
Configure
Time&Date
then
and
Load Factory
XXXXXX
Save config.
No
or
Load config.
No
or
A&B Sensor
NH3
or
Sidebar:
Warning:
Load Factory
&
Load config
Analyzer re-configuration may change sensors,
Reconfiguration and run occurs as soon as either option selected.
Load Factory: ‘LB’ Option
Up to 15 configurations may be included with the Analyzer.
The configurations shipped with the Analyzer may make it easy to add a 2
nd
boiler blowdown,
They may also include complete preset feed programs complete with setpoints.
Recommended:
Turn on Keypad password (Manual Section 5.1).
Log on as the ‘admin’ and modify the password if this Analyzer is Is likely to be accessed by
uniformed users. Do this to prevent accidental or malicious Analyzer reconfiguration.