10.C. Related Products and Instrument Accessories
Product
Size Cat.#
Maxwell
®
16 LEV Cartridge Rack (for use with LEV configuration)
1 each
AS1251
Maxwell
®
16 LEV Magnet (for use with LEV configuration)
1 each
AS1261
Maxwell
®
16 LEV High Strength Magnetic Rod Assembly and Plunger Bar Adaptor
(for use with LEV configuration)
1 each
SP1070
Maxwell
®
16 Cartridge Rack (for use with SEV configuration)
1 each
AS1201
Maxwell
®
16 Magnetic Elution Rack (for use with SEV configuration)
1 each
AS1202
Maxwell
®
16 LEV Magnet
1 each
AS1261
Bar Code Reader for Maxwell
®
16
1 each
AS3200
Thermal, Serial Printer and Cable, Universal Power
1 each
E2821
LEV Plungers
50 each
AS6151
SEV Plungers
50 each
AS5201
LEV Elution Tubes
50 each
AS6201
SEV Elution Tubes
50 each
AS5101
UV Bulb
1 each
SP1080
11. Appendix
11.A. Updating Firmware
As Promega provides new purification kits, new versions of firmware may be required. The firmware version installed
on your instrument can be verified by turning the machine off, then on again. The initial screen will display the version
number of the firmware loaded on the instrument. Please note the firmware version currently installed on your
instrument before contacting Promega or your local Promega representative for new firmware.
Firmware is updated using an SD card reader in the Maxwell
®
16 Forensic Instrument SD slot. Download the updated
firmware as a main.bin file from the Promega web site onto a computer. Place the SD card in the computer’s SD card
slot, and download the updated firmware to the SD card. If you do not have an SD card reader on your computer, use
the enclosed SD card reader to download the updated firmware to the SD card through a USB port on your computer.
Place the SD card containing the firmware into the SD slot on the instrument; make sure the instrument door is closed,
and turn the instrument off and on. The instrument will install the firmware, the Maxwell
®
16 Forensic Instrument
screen will indicate that the firmware is being updated, and the instrument will go through the diagnostic checks.
During the diagnostic checks, the instrument will display the new firmware version. When installation is complete, the
Home
screen will open. If the firmware was loaded successfully, the main.bin file will be deleted from the SD card.