7.A. Changing the Hardware
!
Important.
Do not reconfigure the instrument hardware while samples or reagent cartridges are in place on
the instrument platform. Clean the instrument thoroughly following the instructions in Section 8 before
removing the existing hardware. Wear gloves when changing the hardware to avoid exposure to any potential
contaminants.
Changing from SEV to LEV Hardware
1. Select
Hardware
on the
Maxwell
®
16 Configuration
screen.
2.
Follow the on-screen instructions. First, open the door.
3.
The Magnetic Assembly will move down to allow you to remove the Magnetic Rod Assembly. Do not push down
on the Magnetic Rod Assembly if the instrument is powered on.
4.
Unscrew the three thumbscrews (boxes in Figure 32) on top of the Magnetic Rod Assembly.
5189TB
Figure 32. SEV Magnetic Rod Assembly and thumbscrews.