1.F. Environmental Requirements (Operating, Shipping and Storage Conditions)
Power Requirements: 100–240VAC, 50–60Hz, 2.1A
Temperature: 5–40°C
Humidity: up to 80% relative humidity
The Maxwell
®
16 Forensic Instrument is intended for indoor use only. Wipe up spills immediately. Install the
instrument on a clean, level surface. To avoid shortening the expected lifespan of the instrument, install in a location
that meets the following criteria:
•
Locate on a sturdy, level surface.
•
Avoid dusty areas.
•
Choose a location that has good air circulation and is not exposed to direct sunlight.
•
Avoid noisy electrical power sources (e.g., power generators).
•
Do not install in a location where there is large temperature variability or high humidity.
•
Position the instrument so that it is easy to unplug from the power source.
•
Do not place next to heat sources.
•
Do not use near flammable gases or liquids.
•
Do not place near other electrically sensitive instruments.