The system displays this menu only when you have chosen
the setting REGION from the TIME SCALE menu. Here
you enter the midpoint of the area you want to print out.
Refer to the section above.
(I) Current Scale( internal current
measuring)
Here you set the scaling factor for the internal current
axis. If you choose AUTO, the system automatically sets
the scale so that axis becomes as large as possible. If you
choose not to activate this function, the axis will not be
displayed on the printout.
Option - >
Motion Scale
Here you enter the scaling factor for the motion axis. You
can only use this setting if you have chosen to measure
motion in the ANALOG CHANNEL menu. If you choose
AUTO, the system automatically sets the scale so that axis
becomes as large as possible. If you choose not to activate
this function, the axis will not be displayed on the print-
out.
(X) Current Scale (external current
measuring)
Here you enter the scale for the current axis if you meas-
ure current on the analog channel. You can only use this
setting if you have chosen to measure current in the
ANALOG CHANNEL menu. If you choose AUTO, the
system automatically sets the scale so that axis becomes as
large as possible. If you choose not to activate this
function, the axis will not be displayed on the printout.
Voltage Scale
Here you enter the scale for the voltage axis if you
measure current on the analog channel. You can only use
this setting if you have chosen to measure voltage in the
ANALOG CHANNEL menu. If you choose AUTO, the
system automatically sets the scale so that axis becomes as
large as possible. If you choose not to activate this
function, the axis will not be displayed on the printout.
Option - End
Save in memory
Here you can save the settings in Egil’s memory. You can
store an array of settings in each of the 10 storage areas.
All of the parameters that can be set in Egil’s menus can
be stored in memory with the exception of channel
calibration for internal current axis and motion channel.
However, the actual measurement results are not stored in
memory but are written over by the next measurement or
erased when Egil’s power is switched off.
0
Save setting in memory area 0. The array of settings
stored in memory area 0 are always activated when you
turn on Egil.
1...9
Save settings in memory areas 1 - 9. You choose the
memory area where you will store the settings and press
ENTER to save..
Recall
Here you recall the settings that you have saved in Egil’s
memory.
0...9
Recall settings that have been saved in memory area 0 - 9.
Memory 0 is recalled automatically at power on.
Default
Recall default settings.
The default settings are set at the factory. They are stored
in memory 0 - 9. If you want to use the default settings
every time Egil is switched on do as follows. Choose
RECALL, DEFAULT and SAVE choosing memory 0.
Calibration
Calibration is used for calibrating the internal current axis
channel and the motion channel. Refer to instructions on
calibrating in chapter 11 “Calibration”.
Option - >
Analog channel (option)
Here you can set parameters for measuring motion, among
other things.
The analog channel in Egil is primarily intended for
measuring motion but can also be used to measure current
via a current shunt. It can also be used to measure voltage
directly or via a voltage divider.
The table below shows the possible choices in the
ANALOG CHANNEL menu.
Summary of Contents for EGIL BM-19070
Page 1: ...EGIL Breaker Analyzer User s Manual...
Page 2: ......
Page 7: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 7 E G I L...
Page 9: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 9 E G I L...
Page 15: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 1 5 E G I L...
Page 19: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 1 9 E G I L...
Page 20: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 E R 0 7 A 2 0 E G I L 5 Control panel...
Page 47: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 4 7 E G I L...
Page 49: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 4 9 E G I L...
Page 61: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 6 1 E G I L...
Page 65: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 6 5 E G I L...
Page 75: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 7 5 E G I L...
Page 78: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 E R 0 7 A 7 8 E G I L...
Page 79: ...P r o g r a m m a E l e c t r i c A B Z P B M 0 1 D R 0 7 A 7 9 E G I L...
Page 80: ...Subject to change without notice Printed matter No ZP BM01E R07A Doc BM0087JE...