XTR-PX2-C0006-IN004
7-5
7.
A displacement is automatically calculated and
the result value (OK/NG) is displayed.
When [OK] is displayed, go to “7.3 Adjusting
When [NG] is displayed, take a memo of the
displayed value and go to the next step.
* When Panorama Alignment process takes
time and it does not finish, go to “7.10 When
adjustment work does not finish
8.
Enter the value recorded in your memo in the
column on the right side of the "Colli Offset"
button, and click "Colli Offset".
9.
Wait for that the Colli Offset button returns to
the Enable status.
Go back to step 4 and repeat the Panorama
Alignment process.