XTR-PX2-C0006-IN004 Ceph option
3-3
3.2 Firmware version confirmation
Before starting calibration work, we should confirm the firmware version is matched with other firmware in the
Scanner unit. On the chk1 screen of LCD panel of maintenance mode, we can see firmware versions.
authorized firmware version group ( as of June 1st of 2017 )
System Version
( not displayed )
1.2.1.0
1.3.1.0
Scanner
1.0.8.4846
1.3.1.5010
MTC
1.1.562.0
1.1.593.0
XRC
1.0.37.0
1.0.133.2
CT FPGA
1.0.0.1
1.0.0.1
CT FPDC
1.0.1.0
1.0.1.1
Ceph FPGA
1.0.0.2.
1.0.0.2
Ceph FPDC
1.0.0.3
1.0.0.3
In Ceph unit, there are two firmwares, "Ceph FPGA" and "Ceph FPDC". According to the other firmware
versions, upgrade / downgrade firmware version of "Ceph FPGA" and "Ceph FPDC".
1. Firmware of "Ceph FPGA"
(1) Copy the Ceph firmware from memory device to Ramdisk area
Open \\192.168.5.6\ramdisk with using Explorer.
copy Ceph firmwares to Ramdisk area
(2) Upgrade the firmware with using Maintenance mode
Click "CEPH FPGA" button.
Wait 4 or 5 minutes.
Then there will be message,
"Success: Restart Scanner !"
Turn off and on the power of Scanner
unit.