XTR-PX2-C0006-IN004 Ceph option
4-15
3.
Place the red mark so that the circle of it is
overlapped by the Ear rod ring.
Place the blue mark so that the blue mark
center is overlapped by the pin.
Afterward, click [OK].
4.
Check the
“Result” field, and adjust the
horizontal position of Ear rod ring based on
the displayed adjustment value.
5.
Adjust the horizontal position of Ear rod ring
Loosen the fixing screw,
and rotate the adjustment screw the amount that displayed in the “Result” field. Then
tighten the fixing screw again.
Fixing screw
Adjustment screw
Adjustment direction
DOWN direction