XTR-PX2-C0006-IN004 X-ray door interlock
2
3. Caution
3-1 effective range
The spec sheet shows the below range, we should take it about 10mm distance.
3-2 when the door is made of steel
When the customer uses metallic door in the x-ray room, the effective range will be decreased,
because the magnetic field is affected by the steel door.
Therefore, in case of metallic door, please take care the location of magnet and sensor.
Installation example
of "proximity sensor" and "magnet"
unit : mm