XTR-PX2-C0006-IN004
7-6
7.3
Adjusting Collimator’s position (CT Small Alignment)
1.
Click
“CT Small Alignment”.
2.
Wait for that the status indicator changes from
“green” to “blue”, and from “blue” to “green”.
3.
Press the Exposure Switch to emit X-ray.
4.
A displacement is automatically calculated and
the result value is displayed.
If "OK", go to "7.3 Attaching the Geometry
phantom".
If [NG], go to step 5.