55
Measuring Mode (Bit 9)
Thickness:
If bit 9 is set to 1, then the resulting value is thickness corresponding to the difference of the peak as defined in
bits 3 and 4 and the next peak. This facility is used to request thicknesses directly.
N
OTE
This difference value is not corrected for the index of refraction. Consequently, in order to obtain the real thickness data, one
must divide the result by the refractive index.
Altitude:
If bit 9 is set to 0, then the resulting value is the Altitude corresponding to the peak number defined in bits 3 and 4.
Statistics (Bit 11 to 13)
Statistics are not supported by CLS. Statistics are supported by CLS explorer. Consequently the respective bit
11 to 13 are set to 0. Corresponding to .NET function:
sodx.AltitudePeakSignalOffset1 =
eSodxPeakSignalOffset
.Average;