Model PD6262 Analog Dual-Input Rate/Totalizer Instruction Manual
41
Programming the Rate/Totalizer (
prog
)
It is
very important
to read the following information, before proceeding
to program the meter:
The meter is factory calibrated prior to shipment to read in
milliamps and volts depending on the input selection. The
calibration equipment is certified to NIST standards.
Use
the
Scale
menu to scale the process input (e.g. 4-20 mA). A
calibrated signal source is not needed to scale the meter.
Use
the
Calibrate
menu to apply a signal from a calibrator or a
flowmeter.
The
Program
menu contains the following menus:
1. Scale channel A and B without a signal source
2. Calibrate channel A and B with a calibrated signal source
3. Channel A and B total time base & conversion factors
4. Channel A and B grand total time base & conversion factors
5. Channel A and B reset modes for total & grand total
The process inputs may be calibrated or scaled to any display value
within the range of the meter.
Additional parameters, not needed for most applications, are found in the
Advanced Features
menu; see
Advanced Features Menu
, page 70.