Model PD6262 Analog Dual-Input Rate/Totalizer Instruction Manual
8
SPECIFICATIONS
Except where noted all specifications apply to operation at +25°C.
General
DISPLAY
Upper display: 0.60" (15 mm) high, red LEDs
Lower display: 0.46" (12 mm) high, red LEDs
6 digits each (-99999 to 999999), with lead zero blanking
DISPLAY
INTENSITY
Eight user selectable intensity levels
DISPLAY
UPDATE RATE
5/second (200 ms)
OVERRANGE
Display flashes
999999
UNDERRANGE
Display flashes
-99999
DISPLAY
ASSIGNMENT
The Upper and Lower displays may be assigned to show:
One or more rate channels: Channel A (Ch-A), B
(Ch-B), or C (Ch-C)
Toggle between rate channels: Ch-A & Ch-B, Ch-A
& Ch-C, Ch-B & Ch-C, and Ch-A, Ch-B, & Ch-C
Total or grand total: Ch-A or Ch-B
Rate and total or grand total: Ch-A, Ch-B
Relay set points
Max and/or min values: Ch-A, Ch-B, or Ch-C
Toggle between any rate channel & units
Total and units: Ch-A or Ch-B
Toggle between totals: Ch-A & Ch-B; Ch-A, Ch-B,
and sum of Ch-A and Ch-B
Modbus input
The lower display may also be set to show engineering
units or be off, with no display.
PROGRAMMING
METHODS
Four front panel buttons, digital inputs, PC and MeterView Pro
software, Modbus registers, or cloning using Copy function.
MAX/MIN
DISPLAY
Max/min readings reached by the process are stored until
reset by the user or until power to the meter is cycled.
PASSWORD
Three programmable passwords restrict modification of
programmed settings and two prevent resetting the totals.
Pass 1: Allows use of function keys and digital inputs
Pass 2: Allows use of function keys, digital inputs and editing set/reset points
Pass 3: Restricts all programming, function keys, and digital inputs.
Total: Prevents resetting the total manually
Gtotal: Prevents resetting the grand total manually