Model PD6262 Analog Dual-Input Rate/Totalizer Instruction Manual
84
Display Description
Display Description
Big Hi
Max on big display
(Channel A)
Rst Lo
Reset min
Big Lo
Min on big display
(Channel A)
menu
Menu button
Big HL
Max/min big display
(Channel A)
Right
Right arrow button
Lit Hi
Max on little display
(Channel B)
up
Up arrow button
Lit Lo
Min on little display
(Channel B)
Enter
Enter button
Lit HL
Max/min little display
(Channel B)
Alrm 1
Alarm 1 – 8 (Digital
Output)
Lit HC
Max on little display
(Channel C)
F on 1
Force on relay 1
Lit LC
Min on little display
(Channel C)
F on 2
Force on relay 2
LitHLC
Max/min little display
(Channel C)
F on 3
Force on relay 3
disabl
Disable function key
F on 4
Force on relay 4
Internal Calibration (
ICAL
)
The meter is factory calibrated prior to shipment
to read in milliamps and volts depending on the
input selection. The calibration equipment is
certified to NIST standards.
The use of calibrated signal sources is necessary to perform the internal
calibration of the meter.
Check calibration of the meter at least every 12 months. Each input and
input type must be recalibrated separately.
The
Internal
calibration
menu is part of the
Advanced
Features
menu.
Notes:
1. If meter is in operation and it is intended to accept only one input
type (e.g. 4-20 mA), recalibration of other input is not necessary.
2. Allow the meter to warm up for at least 15 minutes before
performing the internal calibration procedure.