MA-Di
COMPRESSED AIR AT ITS BEST
8.2.4.5
Set standard pressure which the measuring and display values for the
flow rate refer to
rEF.P
Select and set desired standard pressure.
Setting range: 950...1050 hPa in steps of 1 hPa.
8.2.4.6
Set standard temperature which the measuring and display values for
the flow rate refer to
rEF.T
Select and set desired standard temperature.
Setting range: 0...25 °C in steps of 1°C.
8.2.4.7
Set low flow cut-off
LFC
Select and set limit value.
Setting range: 0.1...0.8 Nm³/h in steps of 0.1 Nm³/h.
8.2.5
Service functions
8.2.5.1
Read min / max values for flow
HI
LO
Set
or select,
press
briefly.
HI
= maximum value,
LO
= minimum value
Delete memory
HI
LO
Or
select.
Set
Press and hold until
[----]
is displayed.
Shortly press
Mode / Enter
It is reasonable to delete the memory as soon as the device works
under normal operating conditions for the first time.
8 Progr Parameter setting