18
5.5
Flow direction
For the installation of the measuring armature, the flow direction
must be observed. This is represented by an arrow on the measuring
armature.
The arrow shows the direction of the fluid flow in the piping
line.
5.6
Installation of the sensor in the measuring armature
WARNING
If the
PB+CO®lock
-blind plug is mounted, firstly make sure that the
measuring armature has been closed.
Never remove the PB+CO®lock-blind plug or the sensor with
open measuring armature – this is extremely dangerous.
1. Remove the red transport cover from the sensor element.
2. Mount the sensor into the measuring armature.
Observe the correct installation position of the sensor.
3. Due to the cylinder pin, the sensor can only be mounted in one
position.
4. Fasten the sensor
without any tools
to the measuring armature
using the cap nut.
5. Mount the
PB+CO®lock
into the intended parking station.
6. The mechanical installation of the sensor is now completed.
5 Installation