6
Instructions for Use 2019-03-20
Features
⹅
Indications for Use
The Endostar Provider is a compact, cordless endodontic treatment motorized handpiece for preparation and enlargement of root canals.
It may be connected to the Endostar Navigator (Poldent Co. Ltd.), an apex locator (sold separately).
It can be used to enlarge and prepare root canals, remove of gutta-percha point and softened dentin, and professional mechanical tooth clean-
ing.
Instructions for how to use the device when it is connected to the Endostar Navigator are printed on a blue background.
*
These functions are available only when the device is connected to the Endostar Navigator.
*
These functions are available only when the device is connected to the Endostar Navigator.
⹅
Liquid Crystal Display (LCD)
The LCD is easy to read and shows all settings as well as how the motor is running.
⹅
Controls
< OTR (Optimum Torque Reverse) Mode >
If the file torque is less than the set value, the file will keep rotating in the forward direction.
When the file torque is more than the set value, the file will automatically start rotating 90° in reverse and 180° forward repeatedly. Further-
more, OTR mode can set various motor controls as described below.
Speed :
100, 300, and 500 rpm
Torque Setting
:
0.2, 0.4, 0.6, 0.8, and 1.0 Ncm
Auto Start & Stop
*
:
The file starts when it is inserted in the canal and stops when it is taken out.
Apical Reverse or Stop
*
:
The motor reverses or stops when the tip of the file reaches a preset position inside the canal.
< Normal Mode >
If the file torque is less than the set value, the file will keep rotating in the forward direction. When the file torque is more than the set value, the
file will automatically start rotating in reverse direction.
Furthermore, normal mode can set various motor controls as described below.
Speed :
50, 100, 150, 200, 250, 300, 400, 500, 600, 800, and 1000 rpm
Torque Reverse
:
The motor automatically reverses its rotation if the torque load exceeds the set value to reduces the risk of
jamming.
Slow Down
:
The file slows down as torque increases.
The file slows down as it approaches the apex if the device is connected to the Endostar Navigator.
Forward & Reverse
:
The file may rotate in both forward and reverse directions.
Auto Start & Stop
*
:
The file starts when it is inserted in the canal and stops when it is taken out.
Apical Reverse or Stop
*
:
The motor reverses or stops when the tip of the file reaches a preset position inside the canal.
Apical Torque Reduction
*
:
The automatic torque reverse value is reduced as the file tip approaches the apex.
⹅
Memory
Six combinations of speed, torque, etc. can be memorized.