Hold the File Release Button down and turn the file back and forth
until is lines up with the notch and goes all the way in.
Release the button to secure it.
Lift the electrode up and clip it onto the file.
Replacing the Built-in Electrode with the External File Electrode
• Make sure the file goes all the way in. Give it a light tug to make sure it is held securely.
• Never use stretched or other damaged files.
• Make sure the screw is tight enough. Otherwise, it might come out and be swallowed. Also, measurements might not be accurate.
• Replace the external file electrode if it is worn out as shown in the photo to the left.
• Do not let the cutting part of the file touch the electrode; this will wear it out very quickly.
• Some files cannot be used with this electrode.
• Also the Ni-Ti files noted below cannot be used.
• Those with a file diameter of more than 1.2 mm.
• Those with chuck shanks that are nor perfectly round.
• Gates-Glidden Drills
• Those that have cutting sections with large diameters such as largo burs.
To use these types of files, do not clip on the electrode and use the motor in manual mode.
• Do not use files with shanks larger than the ISO standard.
ISO Standard: Diameter 2.334 to 2.350 mm
• After use, be sure to take the file out.
3
1
2
Always clip the electrode on the file when using it. Otherwise, measure-
ments may not be accurate or rotation may not be properly controlled.
(It may not be possible to measure a canal if blood or some other liquid
overflows the canal or if the canal is completely blocked.)
Worn out