19
1
2
Instructions for Use 2019-03-20
Take off the cap and plug the cable’s white plug all the way
into the motor handpiece.
Plug the cable’s gray plug into the Endostar Navigator.
■
Make sure file electrode is making good contact with the file.
■
Touch the file with the contrary electrode and make sure the meter goes all the way to its end and there are no segments that do not
light up.
Plug: White
Plug: Gray
Do not mix up the cable plugs.
Do not put stress on the transmission cable by twisting, bending, or stretching by wrapping it around the device or Endostar Navigator.
Watch out because the motor might start up when you do this.
• Make sure the plugs go straight in.
• After insertion give plugs a light tug to make sure they are securely connected. Otherwise, data may not be transmitted accurately.
• Do not bump the plugs or drop anything on them when they are plugged in.
• Use only the special cable provided. Other cable could be electrically risky and result in damage or injury. Make sure the file goes all the way in. Give the file a light tug to make
sure it is properly installed.
• Check the meter activity before each patient and do not use the device if all the segments of the display do not light up. This suggests that the meter cannot make an accurate
reading.
* Refer to the user manual for the Endostar Navigator.
Operation Check
Connecting the Transmission Cable