15
Instructions for Use 2019-03-20
Memory Settings: Additional Operation Settings (Normal Mode)
Additional Functions:
Torque Slow Down (TORQ.SL.D.), Linked to canal measurement (APICAL ACT.
**
),
Apical Reverse or Stop (APICAL ACT.
**
), Auto Start and Stop (AUTO STT/STP
**
),
Apical Slow Down (APICAL SL.D.
**
), Apical Torque Reduction (APICAL TRQ.D.
**
)
1. Select a memory number for the standby display; press the Plus or Minus Switch.
2. Hold down the Select Switch for at least 1 second to show the displays for additional operation settings.
3. Press the Select Switch to go from one display to the next.
4. Change the setting; press the Plus or Minus Switch.
* The display will go back to the standby display if 5 seconds (initial setting) elapses without a switch
being pressed.
Torque Slow Down:
When this is set to “ON”, the motor will slow down as the torque load increases.
Standby Display
Apical Torque Reduction
Linked Function
**
:
When this is set to “ON”, Apical Reverse or Stop function be activated.
Apical Reverse or Stop
**
:
The file will reverse or stop when the file tip reaches the Flash Bar.
When Apical Action Function is set to “OFF”, this display will be skipped.
Auto Start & Stop
**
:
When this is set to “ON”, the file starts rotating when it is inserted and stops when it is taken out of the
canal.
Apical Slow Down
**
:
When this is set to “ON”, the file slows down as it approaches the Flash Bar.
Apical Torque Reduction
**
:
When this is set to “ON”, the torque setting that triggers reverse rotation is reduced as the file tip ap-
proaches the apex.
**
These functions are available only when connected to the Endostar Navigator.
Apical Slow Down
Auto Start & Stop
Apical Reverse or Stop
Torque Slow Down
Linked Function
Cannot be used along with Apical Torque Reduction function.
If the Torque Reverse Less (TRL) is turned on, the Torque Slow Down function is disabled.
Cannot be used along with Apical Torque Reduction function.
Cannot be used along with Apical Slow Down or Torque Slow Down functions.
If the Torque Reverse Less (TRL) is set to “ON”, the Apical Torque Reduction function is disabled.