38
Instructions for Use 2019-03-20
Problem
Check Points
Response
Meter is not stable during
use.
• Does the built-in electrode need replacement?
Has it been replaced recently?
• Clean and lubricate the contra angle.
• Take out the built-in electrode and clean it and the rotor
axle with a brush.
• Replace the built-in electrode.
• Is the screw for the built-in electrode loose?
• Tighten the screw.
Motor handpiece will not
go in reverse rotation.
• Is the Reverse Torque set to “TRL” (torque reverse
less)?
• Change the setting to other than “TRL”.
• Is Reverse Torque setting too high?
• Reduce the torque reverse value.
• Is the Apical Action Function set to “OFF”?
• Set it to “ON”.
• Is the Apical Reverse or Stop set to “Stp” (stop)?
• Set it to “Rev” (reverse).
Micromotor changes
speed on its own.
• Is the Apical Slow Down set to “ON”?
• When this is set to “ON”, the motor slows down as the
file approaches the apex.
• Is the Torque Slow Down set to “ON”?
• When this is set to “ON”, the motor slows down as the
torque increases.
p. 15 “Torque Slow Down:”
Device turns off or on its
own.
• Was the device no used for a long time?
• Auto power off was probably activated. Press the Main
Switch to turn the device back on.
• Does “Please Charge” appear in the display?
• Battery must be charged right away.
• This can happen if the battery is very low and a
large load is applied to the file.
• Battery must be charged right away.
Error 01
• Turn the device off and disconnect the transmis-
sion cable; does the same error message appear
when the device is turned back on?
• If the device’s operation is restored by disconnecting the
transmission cable, the problem was only temporary
and there is nothing wrong with it.
• If the same error occurs after disconnecting the trans-
mission cable, there is probably something wrong with
the device.
Contact Poldent Co. Ltd.
• Is there some debris on the connector for the
transmission cable?
• If there is, clean the connector.
• Does the error message appear when the trans-
mission cable is twisted or bent sharply?
• There could be a broken wire inside the cable; replace it
with a new one.
Error 04
• Does this happen repeatedly?
• There may be something wrong with the control board.
(In this case, memory settings cannot be saved, but
they can still be changed even though they will not be
saved.)
Error 06
• Does this happen repeatedly?
• The motor circuits may be malfunctioning. Have the
device repaired.