29
Instructions for Use 2019-03-20
• Do not sterilize the autoclavable components by any method other than autoclaving.
• Components are extremely hot right after autoclaving. Wait for them to cool off before touching.
• Do not leave the components in the autoclave.
Autoclave the contra angle and handpiece rest after use for each patient.
Recommended temperature and time:
In a sterilization pouch, minimum 6 minutes at +134°C (+273.2°F) or minimum 60 minutes at +121°C (+249.8°F).
Minimum drying time after sterilization:
10 minutes.
Thoroughly clean and wash the components before autoclaving. If chemical solutions or foreign debris are not removed, autoclaving could damage or deform the
components.
The sterilization and drying temperatures must not 135°C (+275°F).
No components can be autoclaved other than the contra angle and the handpiece rest.
Take the file out of the contra angle before autoclaving.
Be sure to lubricate the contra angle with the LS OIL before autoclaving it.
Follow file manufacturer’s recommendations for autoclaving files.
Sterilization
Autoclavable Components
• To prevent the spread of serious, life-threatening infections such as HIV and hepatitis B, the components (contra angle, handpiece rest) must be autoclaved after each patient’s treat-
ment has been completed.
Put components in individual autoclave pouches.
Packing