8.6 Trouble-Shooting
The following sample point collection will help you to eliminate eventual faults:
Problem:
Changes in the Setup menu or test menu not possible
Display:
PASSWO FALSE
The password protection is active. Select the menu item
PASSWORD
in the Setup menu.
Press the button OK. Now appears the letter
A
. Select the letter
P
with the arrow key and
confirm with OK. The password protection is now deactivated.
Problem:
Display badly readable
Change the display contrast in the menu item
LCDCTR
.
Problem:
Display
dark
Eventually the temperature in the switching cabinet is too high.
Problem:
Error message over temperature
Check the ventilation (see chap. 4.2) and current setups (see chap. 5.4). Mount a fan.
Problem:
Motor does not run.
1. Check the current setups. Run current and boost current are set to 1 A at delivery.
2. Check the cable and the correct mounting of the connectors.
Problem:
Motor becomes very hot
Check the current settings. The set run or boost current might be too high for the motor
winding. The stop current should be set to above 50% of the run current.
As long as the input Boost is active, the set boost current
always
flows. Even if there are
no Control Pulses, the stop current is
not
activated!
Summary of Contents for 10010115
Page 1: ...Manual 2112 A002 GB CLD Linear Power Stage For Stepper Motor with Plain Text Display...
Page 2: ......
Page 10: ...Manual CLD 1 4 Schematic Diagram Fig 3 Schematic diagram CLD MA 2112 A002 GB 8...
Page 19: ...phytron Fig 5 Mountaing kits 17 MA 2112 A002 GB...
Page 33: ...phytron 8 The Menu Fig 15 Structogram of the menu 31 MA 2112 A002 GB...
Page 49: ...phytron Appendix C Declarations of Conformity 47 MA 2112 A002 GB...
Page 50: ...Manual CLD MA 2112 A002 GB 48...
Page 53: ......
Page 54: ......
Page 55: ......