8 Adjustment of Internal Settings
E-625 Piezo Servo Controller
PZ166E
Version: 2.0.0
49
Requirements
You have read and understood the general notes on startup (p. 23).
Only for the computer-controlled mode:
−
You have read and understood the user manual for the E-816 computer interface
submodule.
−
The required software has been installed on the PC from the PI software CD.
The E-625 was installed (p. 15) properly.
You have installed the positioner in the same way as it will also be used in your
application (corresponding load and orientation).
The E-625 is switched off, i.e. the wide input range power supply is
not
connected to
the power socket over the power cord.
You have opened the E-625 (p. 36).
Tools and Accessories
Only in analog mode:
−
High-precision analog signal source (p. 18).
−
SMB/BNC adapter cable (included in the scope of delivery)
Only in computer-controlled mode:
PC with RS-232 or USB-interface for commanding and installed software from the PI
software CD.
High-precision voltmeter (p. 18):
−
necessary in analog mode
−
not necessary, but helpful, in computer-controlled mode (only E-625.CR)
If the voltmeter is used, an additional SMB/BNC adapter cable is necessary (included in
the scope of delivery).
High-precision, external measuring device (e.g. interferometer) to measure the
displacement of the positioner
Trimmer adjustment tool
Adjusting the Sensor Range in Analog Mode
1.
Select analog mode and open-loop operation (servo mode OFF) using the
Settings
DIP
switch block:
−
1: ON (down)
−
2: OFF (up)
−
3: OFF (up)
−
4: OFF (up)
2.
Connect the signal source to the
ANALOG IN/WTT
SMB socket using an adapter cable.
3.
Connect the voltmeter to the
SENSOR MONITOR
SMB socket using the adapter cable.