8 Adjustment of Internal Settings
42
Version: 2.0.0
PZ166E
E-625 Piezo Servo Controller
INFORMATION
If a E-625.CR is to be interlinked with at least one E-665.CR piezo servo controller for capacitive
sensors:
1.
Configure the E-625.CR as sensor master.
2.
Configure all E-665.CR as sensor slaves (see E-665 user manual).
Requirements
You have read and understood the general notes on the adjustment of settings (p. 35).
All E-625s to be synchronized are disconnected from the power source, i.e., the wide
input range power supplies are
not
connected to power sockets over the power cords.
You have opened the housings for the E-625 to be synchronized (p. 36).
Tools and Accessories
Trimmer adjustment tool, or alternatively a narrow slot screwdriver
Synchronizing the Sensors
Figure 10: Settings for the master/slave switch
1.
Select an E-625 as the sensor master and check the position of the master/slave switch.
The master position (see top figure) is the default setting.
2.
If the master/slave switch on the selected sensor master is not in the master position,
use the trimmer adjustment tool to turn it to the master position.
3.
Set all other E-625 to sensor slave. To do this, turn the master/slave switches to the
slave position (see top figure) using the trimmer adjustment tool.
4.
Close the housings.
5.
Make sure that all devices have the synchronization lines (pin 7 and 8 on the
Network
socket) and at least one ground wire (pin 1, 2 or 5 on the
Network
socket) connected to
the corresponding wire, (for details see pin assignment of the E-625.CN network cable
(p. 73)).