![Physik Instrumente E-625.C0 User Manual Download Page 44](http://html.mh-extra.com/html/physik-instrumente/e-625-c0/e-625-c0_user-manual_1553909044.webp)
8 Adjustment of Internal Settings
38
Version: 2.0.0
PZ166E
E-625 Piezo Servo Controller
Figure 6: E-625.CR: Positions of the components and adjustment elements
8.3.1
Jumper
Jumper X4
The
X4
jumper is located on the main board (p. 37).
Position
Function
1-2
The E-802 servo controller submodule is activated (default settings).
The servo mode can be controlled with switch 3 on the
Settings
DIP switch block
and (only for E-625.CR) over the E-816 computer interface submodule.
The slew rate limiter for the piezo voltage and the notch filter remain active even
when servo mode is switched off.
2-3
The E-802 servo controller submodule is bypassed, independent of all other
settings.
The slew rate limiter for the piezo voltage and the notch filter are inactive.