8 Adjustment of Internal Settings
E-625 Piezo Servo Controller
PZ166E
Version: 2.0.0
47
Figure 13: Closed-loop operation, P-term set too large (strong overshoot)
Figure 14: Closed-loop operation, conservative setting of P-term and I-term, stable system with
relatively long settling time
8.6
Calibrating the Displacement of the Positioner
A recalibration of the displacement is only necessary when the positioner (or parts of it) or the
E-625 of a calibrated system was replaced. A calibrated, external measuring device is used to
recalibrate the position sensor back to the accuracy specified in the original measurement
protocol.