![Physik Instrumente E-625.C0 User Manual Download Page 23](http://html.mh-extra.com/html/physik-instrumente/e-625-c0/e-625-c0_user-manual_1553909023.webp)
5 Installation
E-625 Piezo Servo Controller
PZ166E
Version: 2.0.0
17
Tools and Accessories
15 V wide-range-input power supply included (for line voltages between 100 and 240
volts alternating current voltage at 50 or 60 Hz)
Alternative: Sufficiently rated power adapter
Barrel-to-Switchcraft adapter for the power supply (K050B0002, included in the scope
of delivery)
Alternative: Sufficiently sized adapter
Power cord included
Alternative: Sufficiently sized power cord
Connecting the Power Supply to the E-625
1.
Connect the adapter's Switchcraft connector (f) to the E-625's
DC IN 12–30 V
Switchcraft panel plug.
2.
Connect the adapter's barrel connector to the power supply's barrel connector socket.
3.
Connect the power cord to the power supply.
5.5
Connect a Positioner to the E-625
Requirements
The E-625 is switched off, i.e. the wide-range-input power supply is
not
connected to
the power socket over the power cord.
Tools and Accessories
The positioner with which the E-625 was calibrated
Connecting the Positioner
Connect the positioner to the
PZT & SENSOR
socket.
If your system was calibrated by PI, the piezo servo controller and the positioner must
not be replaced. Take note of the assignment indicated by the serial numbers on the
calibration label on the piezo servo controller.